Measurement uncertainty of rotary solder paste viscometer in instrument calibration - Database & Sql Blog Articles

Photocoupler

Solder paste, also known as solder paste, is a technology associated with surface filming

(Surface Mount Assembly,

Abbreviation

SMT)

And the new welding materials produced. Its main function is to form solder joints in components and boards to connect metals. Solder paste is a non-Newtonian fluid whose composition and quality can be reflected by viscosity. In solder paste printing, different types of printing techniques and

PCB

The thickness of the printed board on the board is highly dependent on the optimum solder paste viscosity value, so accurate solder paste viscosity values ​​are important in solder paste printing. The solder paste viscometer is mainly used to measure the viscosity value of the solder paste liquid. It uses a rotary pump sensor and a built-in coaxial double cylinder. It is a liquid viscosity physical analysis instrument. It measures viscosity in industrial production. To monitor the composition and quality of the substance. Therefore, solder paste viscometers require periodic calibration to ensure the accuracy and stability of their measurements. For the calibration of solder paste viscometer, refer to the national instrument metrology verification procedure.

JJG1002-2005

Rotational viscometer verification procedures.

1

Instrument measurement method

1.1

Measurement basis

according to"

JJG1002-2005

Rotational viscometer verification procedure, solder paste viscometer

Instrument measurement

The relative method is used, that is, the standard solution is measured using a viscometer.

(

The main component is silicone oil

)

A method of comparing viscosity values ​​to standard values.

1.2

Viscosity measurement

Since the change in temperature has a great influence on the viscosity of the solution, it is in the rotary viscometer.

Instrument calibration

The indoor temperature and temperature conditions during the process are

20

±

1

°C, relative humidity is not greater than

65%

. Before performing the test, the inner cylinder, outer cylinder and viscosity cup of the solder paste viscometer rotor should be ultrasonically cleaned and ensured to be dry and clean before use.

The solder paste viscometer is placed on a horizontal work surface and the glass containing the standard viscosity liquid is placed in the cup. Turn on the viscometer power switch and adjust its indication value in case of idling

000.0

By arranging the up and down movement buttons, the rotor is immersed in the standard viscosity liquid so that the liquid surface is at the sensor discharge port.

5~10mm

At the speed, the speed is set to

10rpm

, the temperature is set to

20

°C. The temperature value of the instrument dial is

20

Press the start-up operation at °C until the reading is stable and record the measured value. Measure under the same conditions

10

And record the measurement results.

2

, uncertainty sources and analysis

2.1

mathematical model

y(

η

)=f(Ka)

Η——the power of the standard liquid at a specific temperature;

K

- viscometer constant;

A

- Deflection angle.

Since the individual inputs are independent of each other, the relative standard uncertainty is:

2.2

Uncertain source analysis

The factors affecting the measurement results include: the uncertainty caused by the repetitive measurement; the uncertainty caused by the indication error of the platinum resistance temperature measurement; the uncertainty caused by the resolution of the instrument display; the uncertainty caused by the standard solution setting, such as Figure

1

Shown as follows:

2.3

Measurement uncertainty assessment

2.3.1

Measuring the uncertainty caused by repeatability

u r


Viscosity value

13.8Pa

·

s -1

The standard viscosity liquid is measured ten times, and the measured results and environmental recording conditions are as follows.

1

:

Calculate the average based on the experimental results measured in the table

x

for

13.78Pa

·

s -1

The residuals of each measurement are:

-0.12Pa

·

s -1

,

0.12Pa

·

s -1

,

0.02Pa

·

s -1

,

0.08Pa

·

s -1

,

0.02Pa

·

s -1

,

0.08Pa

·

s -1

,

-0.18Pa

·

s -1

,

-0.18Pa

·

s -1

,

-0.08Pa

·

s -1

,

-0.02Pa

·

s -1

, using the Bezier formula to calculate the standard deviation:


2.3.2

Uncertainty caused by thermometer measurement error

u t

A platinum resistance thermometer is used in the calibration process of the instrument, and the index is

0.1

°C, because the limit error cannot exceed the minimum index

3

Times, temperature change

1

The change in viscosity caused by °C is average

3%

, calculation by repeatability results

13.78Pa

·

s -1

×

3%=0.4134Pa

·

s -1

.

JJG1002-2005

The temperature fluctuations required in the test do not exceed

0.1

°C, the resulting maximum viscosity change is

0.4134

×

0.1=0.04134Pa

·

s -1

.

,

uncertainty

u t

for:

2.3.3

Calculation shows the uncertainty introduced by resolution

u s

The numerical value of the scale value is

d=0.1Pa

·

s -1

, display resolution is

1d

Since the uncertainty caused by such errors is the quantitative uncertainty of the electronic counter, it is subject to uniform distribution.

,

uncertainty

u s

for:

2.3.4

Standard viscosity liquid setting uncertainty

u l

The viscosity of the standard viscosity liquid is generally half a year, and the viscosity value will change slightly during the period from the preparation to the use.

JJG1002-2005

Permissible error requirements are given in the Rotational Viscometer Calibration Procedure

0.1%

~

0.5%

With maximum tolerance

0.5%

Calculate, calculate the uncertainty introduced by the standard solution according to the uniform distribution

u 1

:

2.5

Synthetic standard uncertainty

u c

And extended uncertainty

3

Conclusion

Solder paste viscometer

Instrument calibration

The process is affected by many factors. According to the calculation result of uncertainty, the influence of relevant factors can be minimized, and the quality of instrument calibration work can be improved.

Acer Adapter

High efficient charging speed for Acer laptop, stable current outlet can offer power for the laptop at the same time charge the laptop battery. The best choice for your replacement adapter. The DC connector is 5.5*1.7mm or 3.3*1.0mm. We can meet your specific requirement of the products, like label design. The plug type is US/UK/AU/EU. The material of this product is PC+ABS. All condition of our product is 100% brand new.

Our products built with input/output overvoltage protection, input/output overcurrent protection, over temperature protection, over power protection and short circuit protection. You can send more details of this product, so that we can offer best service to you!

Laptop Adapter For Acer,Charger For Acer,Acer Laptop Adapter ,Ac Adapter For Acer

Shenzhen Waweis Technology Co., Ltd. , https://www.waweispowerasdapter.com

This entry was posted in on